Manual Temperature Setting; Ph Simulation; Redox Simulation - ProMinent DULCOMETER Pt 1000 Instrucciones De Servicio

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Operation
6.4.5
Press the key "CAL" while switching on the meter to reload the default
calibration settings (zero point: pH 7.00; slope: 100 %).
6.5
Setting the temperature value for manual temperature compensation of the pH
measurement without a Pt 1000 gauge.
Press the "SIM" key to select the "manual temperature compensation"
(SIM Pt 1000) operating state.
LCD markers "SIM" and "°C" are activated.
Enter the temperature value, between 0 and +100 °C, resolution 0.1 °C,
using the "UP" or "DOWN" keys as necessary.
6.6
This feature is intended for checking other pH meters and cables. It can only
be selected if no Pt 1000 electrode is connected to the "temp/sim" connector.
The output setting time is max. 60 seconds.
Press the key "SIM" to select the SIM pH operating state.
LCD markers "SIM" and "pH" are activated.
Enter the pH value to be simulated, between 0 and 14 pH, resolution
0.01 pH, using the "UP" or "DOWN" keys as necessary.
The signal is output at the tem/sim connector; output impedance: 100 MΩ.
6.7
This feature is intended for checking other redox meters and cables. It can only
be selected if no Pt 1000 electrode is connected to the temp/sim" connector.
The output setting time is max. 60 seconds.
Press the key "SIM" to select the SIM redox operating state.
LCD markers "SIM" and "mV" are activated.
Enter the mV value to be simulated, between -1300 and 1300 mV,
resolution 1 mV, using the "UP" or "DOWN" keys as necessary.
The signal is output at the "temp/sim" connector.
output impedance: 100 MΩ
accuracy: ±3 mV.