Replacement Of Fuses - Bien-Air Dental Chiropro Plus 3rd Gen Instrucciones De Uso

Ocultar thumbs Ver también para Chiropro Plus 3rd Gen:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 91
FIG. 2

12.4 Replacement of fuses

A. Switch off the Chiropro Plus 3
B. Disconnect the mains cable.
CAUTION
The power cable must be disconnected at least 10 seconds before
opening the fusebox.
FIG. 2
C. Remove the fuse box (1) with a flat screwdriver.
FIG. 3
D. Replace the fuses (2) by the new ones and put the fuse box
back (1) in place.
CAUTION
Only use fuses T4.0AH 250 VAC REF 1307312-010.
1
1
rd
Gen unit.
2
FIG. 3
NOTES
1
Bien-Air Dental SA recommends the user to have its dynamic
instruments regularly checked or inspected.
2
The knob is hold magnetically. There is no need to preserve its
angular position when removing it or putting it back in place.
25

Publicidad

Tabla de contenido
loading

Tabla de contenido