FF descriptor
APPLICATION_TY-
PE_SEL
TEMP_COMP_UNIT
DELTA_I
GAUGE_TEMPERA-
TURE
DECAY_COMPENSA-
TION_FACTOR
PMT_VOLTAGE_CALI-
BRATION
CORRELATION_CO-
EFF
ERROR_TEXT
PMT_VOLTAGE_AC-
TUAL
STANDARDIZATION_
FACTOR
SERIALNUMBER
NAMUR_STATE
NULL_COUNT_RATE
COUNT_RATE_PMT
ADJ_DENSITY_ABS_
COEFF
DEV_SW_VER_ASCII
POINT_LEVEL_AD-
JUST_MODE
RELAY_VALUE_SEL
DIGITAL_IN
DIGITAL_IN_BOOL
ANALOG_IN
FF_CHANNEL_AVAI-
LABLE
FF_CHANNEL_USED
Mapping of Process Value Status
Hex
Quality
0x00
Bad
0x01
Bad
SOLITRAC 31 • Foundation Fieldbus
Rel. In-
Description
dex
38
Selected application
39
Selected unit code for process temperature
40
Calculated percent delta I
41
Electronics temperature
42
Factor for the decay compensation
43
Photomultiplier voltage on delivery
44
Correlation coefficient for linearizer table
45
Error text
46
Current photomultiplier voltage
47
Factor for the real value correction
48
Serial number
49
NAMUR state
50
Zero count rate
51
Pulse rate photomultiplier (raw values)
52
Process absorption coefficient
53
Software version
54
Point level adjustment mode
55
Relay basic value
56
Frequency of digital input
57
State of digital input
58
Current on analog input
59
Available channels
60
Used channels
Sub-Status
Non-specific
Configuration Error
Condition
Unexpected error
It was tried to set a wrong unit with FF interface
On user error codes: 16, 17, 25, 52, 57, 66, 72,
117, 120
11 Anexo
Unit
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
77