Halyard ON-Q Instrucciones De Uso página 8

Ocultar thumbs Ver también para ON-Q:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 25
Warning
Do not rely on the SELECT-A-FLOW* dial key
:
or tie-wrap to prevent patient tampering.
SELECT-A-FLOW* device is available in two flow rate ranges
and distinquished by color on the face of device:
Blue (1-7 ml/hr)
Flow rate: 1, 2, 3, 4, 5, 6, 7 ml/hr
Purple (2-14 ml/hr)
Flow rate: 2, 4, 6, 8, 10, 12, 14 ml/hr
Instructions for Use
Use Aseptic Technique
Filling The ON-Q* Pump: (Figure 4)
Note: Follow hospital protocols and applicable regulations for
filling pump.
1. Close clamp.
2. Uncap the fill port.
3. Attach filled syringe to fill port. Invert pump as shown.
4. Grasp syringe with both hands.
5. Push down on plunger continuously until volume is
dispensed. Do not handle pump while filling, as the
syringe tip may break. Repeat as necessary. Syringe
accuracy is ±4%.
Note: Filling Extension Sets are provided with larger
pumps (see product insert).
6. Remove filling device from fill port
7. Replace fill port cap. Label with appropriate
pharmaceutical and patient information.
Figure 4
6
Fill Volume
Caution: Do not underfill pump. Underfilling the pump
may significantly increase the flow rate. Do not exceed
maximum fill volume. (Table 1)
Table 1: Fill Volume (ml)
100
Fill Volume
125
Max Fill Volume
Retained Volume
Warning: Flow rate is adjustable. Medication dosage
should be based on maximum flow rate. Pump is pre-set
at 7 ml/hr or 14 ml/hr (depending on model).
Table 2: 24-Hour Dosing Reference
SELECT-A-FLOW*
Model
7 ml/hr Max Flow Rate
14 ml/hr Max Flow Rate
Formula:
ml/hr x % drug concentration x 10 x 24 hr = 24 hour dose (mg)
Caution: Calculations based on the labeled flow rate.
Flow rate accuracy varies.
Priming the Administration Set
Use Aseptic Technique
Note: SELECT-A-FLOW* device is packaged with the flow rate
at the highest setting to minimize priming time.
1. Open the plastic cover of the SELECT-A-FLOW* device
(Figure 5)
2. To begin priming the SELECT-A-FLOW* device, ensure
highest flow rate setting is selected.
• Make sure the selected flow rate is aligned within the
window below the ml/hr x mark.
• The tactile feel will allow the user to ensure selected
flow rate is set.
3. Open the clamp and remove tubing cap to begin priming.
200
270
400
335
335
550
≤5
≤11
≤11
≤16
Drug Concentration (mg)
0.2%
0.25%
336
420
672
840
600
750
≤23
0.5%
840
1680

Publicidad

Tabla de contenido
loading

Productos relacionados para Halyard ON-Q

Tabla de contenido