Vibration Isolator Installation; Typical Layout Diagram; Connection A - High Vacuum Flange - Varian Turbo-V70LP Manual De Instrucciones

Tabla de contenido

Publicidad

Screw the flange mod. 969-9109 on the pump, tak-
ing care of the o-ring right position.
Assemble the seal ring and lock the vent device in
position using the KF klamp.

VIBRATION ISOLATOR INSTALLATION

A vibration damper model 969-9340 for ISO 63
inlet flange version pump is available as accesso-
ries.
It typically reduces the vibration transmitted from
the Turbo-V 70LP pump to the system by a factor
of 20.
Please refer to the relevant instruction manual.

TYPICAL LAYOUT DIAGRAM

1.
Turbo-V controller
2.
Vent valve
3.
Vacuum pump shut-off valve (optional)
4.
System vent valve (optional)
5.
Vacuum chamber
6.
Ionization gauge
7.
Fore-vacuum pump connecting flange
8.
Oil mist eliminator
9.
Fore-vacuum pump with internal one-way
valve
10. Fore-vacuum pump control relay
11. Connection for water cooling
12. Roughing line with valve (optional)
13. Turbopump

Connection A - HIGH VACUUM FLANGE

To connect the Turbo-V70LP pump to the ISO inlet
flange, remove the outer ring and position the cen-
tering ring as shown in the figure.
Then fix the two flanges with the clamps or claws
as shown in the figure.
61
TECHNICAL INFORMATION
87-900-862-01(G)

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

969-9365969-9366969-9367969-9368

Tabla de contenido