plugging the device in, ensure that the voltage from the outlet is within the permissible range
(see technical data) 3. After charging has been completed, please remove the batteries from
the device and unplug the charging station's power cord from the wall outlet. 4. Only use in
closed and dry rooms.
h
Guarantee: When used under normal circumstances and with attention to these instructions
(the User's Manual) we guarantee this product for two years after the date of purchase. Should
you need further information or should your instrument require repair, please contact your
dealer.
Operation: After the transformer plug has been plugged into the outlet, the charging station
is ready for use. (The green LED display will read: "stand by".) When the rechargeable handles
d
are inserted into the station, the LEDs on the respective charging slots will display the charging
e
progress (red = charging, green = fully charged, blinking red = error). After the target charging
n
voltage has been reached, the LED changes from red to green, i.e. the rechargeable handle has
been fully charged. Charging begins automatically as soon as the handle is inserted into the
charging slot. The charging slot can be adapted as needed to accommodate the different types
V
of charging handles. The rechargeable battery can be removed from the handle / replaced by
unscrewing the cap over the battery compartment.

Performance characteristics:

- Input voltage .....................................100V-240 V+/-10%, 50-60 Hz
- Max. power .......................................1.2 A
- Efficiency factor .................................90-93%
y
- Power factor ......................................>0.99
- Charging cells ....................................NiMH und Li-Ion
- Nr. of cells ..........................................NiMH 2-3 cells, Li-Ion 1 cell
- Chargeable capacity ..........................min. 250 mAh, max. 4500 mAh
- Input voltage charging station..........9 VDC +/-10%
- Power consumption in standby ......< 0.5 W
- Rated power consumption ................approx. 4.5 W
- Output voltage in standby ...............< 9VDC
9