With the RS 232/422/485 connected, a bi-directional
communication is established by selecting HOST. Data are
sent to an external computer every time the external
computer asks for the values.
The data available are:
•
Speed adjust
•
Pump/controller operating condition
•
Cycle time
•
Pump life
•
Pump temperature
•
Pump current
•
Pump voltage
•
Controller output frequency
•
Cycle number
•
R1 condition
•
R2 condition
•
Life time and cycle # zeroing
•
Configuration parameter readings
•
Configuration parameter setting
Note that the new input value are put in effect only at the
next STOP/START of the pump.
If PRINT is selected and a printer is connected on RS
232/422
line,
an
unidirectional
established and every minute the data are sent to the
printer, even if the pump is not running.
The set of data available are:
•
Pump speed KRPM
•
Pump temperature
•
Pump current A
•
Pump power W
•
R1 condition
•
R2 condition
The controller is factory-set to HOST.
communication
is
67
TECHNICAL INFORMATION
•
Confirm the selection by pressing the PUMP
CURRENT push-button.
•
In this way you enter into an operating phase named
"Monitor Relay Programming" described in the
following paragraph.
Monitor Relay Programming
•
The display shows:
S P E E D
T H R E S H O L D
S E L E C T
I O N : X X K R P M
where: XXKRPM = is the switch point of relay R1 at the
preset turbopump speed, adjustable from 00 to 99 KRPM.
The speed threshold will condition the R1 and R2
operation (see the following cycle diagram] and it is
factory-set to 57 KRPM. Press the CYCLE NUMBER push-
button to select the first number.
S P E E D
T H R E S H O L D
S E L E C T
I O N : X X K R P M
87-900-893-01 (C)