Disconnect Acp Tubing Set; End Of Procedure; Compatibility Footswitch - Acp System; Trouble Shooting - MORIA 65112 Manual Del Usuario

Sistema de acp
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 26
E. DISCONNECT ACP TUBING SET
Steps
What to do
15

Disconnect ACP tubing set.

F. END OF PROCEDURE
Steps
What to do
16
Switch off ACP system
turning button "O/I" to
"0".
X. COMPATIBILITY FOOTSWITCH – ACP SYSTEM
ACP footswitch (#19189) must be used only with ACP system (#19188).
XI. TROUBLE SHOOTING
A. ACP SYSTEM
Message
Reason

ACP system

No power.
doesn't switch
Failure of the power supply, the
on
supply cord connection or the
fuse.
ACP system
Loss of power.
switches off
Failure of the power supply, the
during operation
supply cord connection or the
fuse.
B. DIGITAL DISPLAY
Message
Reason
Same displayed
Failure of display card
value whatever
the position of
the turning knob
or 888 display
or lines display
or loss of digital
display
C. PRESSURE SETTING
Message
Reason
The pressure
No connection of ACP tubing set
doesn't increase
or leakage
nor decrease
Failure of internal components
with the turning
knob
8
MORIA
#65112-D-12.2017
IMPORTANT WARNINGS
Finger tight only.
Discard ACP tubing set after each procedure in an appropriate container.
IMPORTANT WARNINGS
Visually check that:
• the yellow LED is OFF.
• the green LED is OFF.
• the display is OFF.
Intervention
Check the connection of the main ACP power supply, the green LED of
power supply must be ON.
If not, unplug the system and check the integrity of the fuse and replace
it if damaged.
If unable to solve the problem: do not use ACP system and contact your
distributor or MORIA to have it fixed.
Check the connection of the main ACP power supply, the green LED of
power supply must be ON.
If not, unplug the system and check the integrity of the fuse and replace
it if damaged.
If able to solve the problem; check if the corneal tissue has not been
damaged and start again the different steps described in paragraph VIII.
If unable to solve the problem: do not use ACP system and contact your
distributor or MORIA to have it fixed.
Intervention
Do not use ACP system and contact your distributor or MORIA to have it
fixed.
Intervention
Check the integrity of ACP tubing set. If leakage is present, change it.
Check the sealing of the artificial chamber too.
Do not use ACP system and contact your distributor or MORIA to have it
fixed.
Related
pictures
A2
Related
pictures
B2
A3-A5-A6

Publicidad

Tabla de contenido
loading

Tabla de contenido