Power frequency
(50 or 60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
• Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should
only be used in such environments:
Immunity test
IEC 60601
test level
Conducted RF
150 kHz to
IEC 61000-4-6
80 MHz
3 Vrms to
6 Vrms
(in ISM and
amateur ra-
dio bands)
80% Am at
1 kHz
Radiated RF
10 Vrms,
IEC 61000-4-3
80% Am at
1 kHz
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
30 A/m
50 Hz/60 Hz
Compliance
level
150 kHz to
80 MHz
3 Vrms to
6 Vrms
(in ISM and
amateur
radio bands)
80% Am at
1 kHz
10 Vrms,
80% Am at
1 kHz
30 A/m
50 Hz/60 Hz
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the device, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
d = 0.35
d = 1.2
80 MHz to
where P is the maximum
800 MHz
output power rating of the
d = 1.2
transmitter in watts (W)
800 MHz to
according to the transmitter
2.7 GHz
manufacturer, and d is the
d = 2.3
recommended separation
distance in metres (m). Field
strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, a should be less than
the compliance level in each
frequency range.b
Interference may occur in the
vicinity of equipment marked
with the following symbol:
26
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.