• The electrodes are too old and need
12
to be replaced.
Cannot change program
Check if the code lock is activated.
Follow the instructions under the
Code Lock heading.
TECHNICAL DATA
Number of channels:
Stimulation forms:
Constant current:
Amplitude (strength of current):
Wave form:
Carrier frequency:
Treatment frequency:
Vector scan:
l r.m.s. max/channel:
Power source:
Outer dimensions:
Weight:
Environment for storage,
use and transport:
EXPLANATION OF SYMBOLS
Read the operating instructions prior to use.
Meets the requirements of MDD 93/42/EEC. Reported to SEMKO AB.
0413
Warranty: The stimulator has a two-year warranty against defects, on condition that it is used correctly.
2470-200509_CEFAR_IFC_manual_inlaga_outline.indd
12
You should turn in your CEFAR IFC
for service once a year. Service should be
done by Cefar or one of Cefar's distri-
butors.
2 (simultaneous)
Interference
Premodulated
Up to a resistance of 500 ohms.
0–50 mA in steps of 0.5 mA.
Symmetric biphasic square pulse.
4000 Hz
1–150 Hz
0–100 %
50 mA
Two 1.5-V non-rechargeable AA batteries,
Two 1.2-V rechargeable AA batteries,
AC/DC Adapter 3V Friwo FW 7333M/03 (Cefar
article no. 1554). Note: Do not use the CEFAR
IFC with any other adapter.
65 x 120 x 30 mm
about 160 g.
Temperature 10°C–40°C.
Humidity 30%–75%.
Air pressure 700 hPa–1,060 hPa.
Patient body part, type BF – Body Floating.
2005-09-14, 15:18