COMPLIANCE

The CEFAR IFC makes it easy to
monitor how much the patient uses the
stimulator.
1. Press both
button simultaneously, with the sti-
mulator switched off. The number of
treatments is displayed first.
2. Press P. The average treatment time
is displayed in hours (h) and minutes
(min).
3. Press P. The total treatment time is
displayed in days (d), hours (h) and
minutes (min).
4. Press and hold C for three seconds to
reset.
5. Press

CODE LOCK

The CEFAR IFC lets the trial staff
lock the stimulator into one specific
program to ensure that the patient uses
the right treatment.
Activating the code lock
1. Select the desired program.
2. Switch the stimulator off with
3. Press both
button simultaneously.
4. Select "On" using P. Now the code
lock is activated. Switch the stimula-
tor off with
2470-200509_CEFAR_IFC_manual_inlaga_outline.indd
buttons and the
to close.
buttons and the
.
9
Deactivating the code lock
1. Switch the stimulator off with
2. Press both
button simultaneously.
3. Select "OFF" using P. Now the code
lock is deactivated. Switch the stimu-
lator off with

BATTERIES

A battery indicator on the display of
the stimulator lights when it is almost
time to change the batteries. You can
keep using the same batteries as long
as the stimulation feels the same. You
know it is time to replace the batteries
when the stimulation is weaker than
usual or the stimulator stops working.
If you use rechargeable batteries, it
is time to recharge when the battery
indicator glows or when the stimu-
lator stops working. Note: Never try
to charge non-rechargeable batteries.
They may explode.
To change batteries: First, switch off
the stimulator. Remove the battery
cover by pulling it down and then
lifting it. Remove the batteries. Make
.
sure that the new batteries are installed
in the correct position (match up the
+ and – symbols). Always dispose of
spent batteries properly.
The CEFAR IFC can also run on the
current from a regular wall outlet.
Connect the adaptor cord to the jack
.
buttons and the
.
2005-09-14, 15:18