10.2. How to clean the strip holder

Wipe off urine residues from the strip holder with a lint-free cloth after each mea-
surement. This prevents carry-over and drying of urine residues.
To remove the strip holder for more comprehensive cleaning, make sure to turn
the instrument off. Pull the strip holder from the metal sled below (Pic. 16). Wipe
down the strip holder with disinfection wipes. When removed from the instru-
ment, you may also use water with a soft brush and/or liquid disinfectants.
After cleaning, insert the holder on the metal sled (Pic. 17). The rectangular
notches of the transport mechanism and the strip holder must be placed on top
of each other (Pic. 18). Use reasonable force to push the strip holder completely
onto the metal sled. Doing so the metal slit and the strip holder will be pushed
back completely into the housing. If you get an error message after turning the
instrument on: Turn the instrument off and push the strip holder again on the
metal slit with more force.
Pic. 16: test sled
(bottom view)
ATTENTION:
Ensure that the strip holder is completely clean and dry before inserting.
96
®
Relax
Pic. 17: Notch A
Manual URYXXON
Pic. 18: Notch B
®
Relax EN, V6.00 / 11.20