HC151 PHRX
Digital multifunction solenoid metering pump
4.1.7 SETUP – PROBE CALIBRATION
The probe calibration menu allows you to calibrate the pH probe or Redox probe.
The menu consists:
4.1.7.1 SETUP – PH PROBECALIBRATION
ENTER
If calibration was performed in wrong way, MIS. PH will blink and read value will be 0. You can't set Ph
working mode in working mode menu. Set YES in REST.PAR.PH menu to resolve it.
ADSP9500092
CALIBR. PROBE
ENTER
CALIBR. PROBE
PH/
RX/
Press and hold the ESC key to return to the previous menu.
CALIBR. PROBE
PH/
RX/
ENTER
SOL.
PH 7.00
MIS.
PH 7.00
ENTER
SOL.
PH 4.01
MIS.
PH 4.01
ENTER
PH QUALITY
<N> %
ENTER
REST. PAR. PH
NO/YES
Press and hold the ESC key to return to the previous menu.
14/04/2015 Rev. 1.0
Use Up & Down
calibration or Redox probe calibration.
Use Enter to confirm.
PH
The first step to be performed, is the zero
calibration of the probe, by immersing the same in
a solution of pH 7.
The second step to be performed, is the calibration
of the probe gain, immersing the same in a solution
of known value (eg. 4.01 ph). In the event that the
measured value does not correspond to the value of
the buffer solution, with the Up and Down buttons
you can change the value of the solution.
In automatic mode the pump will calculate the
quality of the probe.
Use Up & down to decide whether to restore the
values of the previous calibration or whether to
store the calibration values just made.
ENGLISH
to select PH probe
42/123