The acoustics status will be displayed.
2. Press FORWARD
The wanted acoustic status will be displayed.
3. Press ENTER to confirm.

RESET OF PROGRAMMING

This menu allows to come back to the initial configuration -factory- by default, resetting everything programmed up till
now.
1. Press PROGRAMMING. Insert your Engineer Code again and ALL parameters will come back to the initial
configuration
"OK" will be displayed and a confirmation acoustic will go off.

ERASING EVENTS

This menu allows to erase all events memorised.
1. Press PROGRAMMING and you will erase all events memorised.
"OK" will be displayed and a confirmation acoustic will go off.
Once you have erased all events from the memory, it will just remain registered the ERASING EVENTS operation with the
date and time in which it took place.

CALIBRATION

All detectors manufactured by DURAN ELECTRONICA have been calibrated in our laboratories, for this reason, they
do not need to be calibrated when being installed.
Please remember that an overhaul of the detectors needs to be effected, at least, once in a year.
Before being calibrated, the input of gas of all detectors must be covered for at least 30 min -except the explosive gases
and O
detectors-.
2
1. Press PROGRAMMING.
The Module line will look for the number and type of detector that are connected, and "Detector Search"
will be displayed. Next "Calibration" and the type of gas will be displayed on the top line; and in the bottom
line the detector nº and its concentration will be displayed.
2. Once the first detector is calibrated press FORWARD to go on to the next one.
3. Press ENTER to exit.