The device might be locked into a pre-defined program selected by your Clinician. In this case, you cannot change the
program selection and the Program Selection buttons are inactive. A corresponding text will appear below the program
number.
4.
The sequence indicator will indicate the number of sequences for each program. The program timer indicates the total
duration of the program. Once you have selected the desired program, simply increase the intensities of the channels you
want using the Increase Intensity buttons. For intensity strength information, refer to 6.6.
The intensity of each channel will be displayed both numerically and with a bar graph. Note that
5.
intensities of Disuse Atrophy (P1) and Reinforcement (P2) programs are expressed in Energy
levels (max 520) while intensities of TENS (P3) and Edema (P4) programs are expressed in mA
(max 100 mA).
7.4.4 Pause function
During a treatment, you can pause the treatment at any time by pressing the ON/OFF
intensities will drop to 0, the timer will stop, and the timer indicator will blink.
To resume the treatment, increase the intensities to the desired level. This will automatically resume the timer.
When you are in the Pause mode, you can also switch off the unit by pressing the ON/OFF
7.4.5 End of treatment
For all treatments except P3 (TENS), the timer will automatically finish the treatment. When the treatment is finished,
intensities are automatically reduced to 0 and the timer indicator will flash with 00:00.
For the TENS (P3) program, the timer is unlimited, meaning that it does not automatically shut off after a set time. Stop the
treatment after the time recommended by your healthcare provider by pressing the ON/OFF
• Turn off the device by pressing the ON/OFF
Disconnect the electrodes and the lead wires as indicated in section 7.3. Electrodes no longer fit for use can be disposed of
with the normal domestic waste.
7.4.6 Use of the hand switch
The hand switch connects to the unit on the dedicated port:
The hand switch can be used to manually change the cycles of contractions/active rest or turn the stimulation On and OFF.
• In a constant mode (P3, P4, warm up and cool down of P1 and P2), pushing the button will activate the stimulation while
releasing it will bring all intensities to 0.
• In Work mode, pushing and releasing the button, respectively, will toggle between contraction and active-rest phases.
22
button.
optional)
(
button once. In this case, both
button a second time.
button.