Empi Phoenix is intended for use in the electromagnetic environment specified below. The customer or user of the Empi
Phoenix should assure that it is used in such an environment
Immunity test
Test level IEC 60601
Electrostatic
discharge (DES)
±6 kV contact
CEI 61000-4-2
±8 kV air
±2 kV for power
Electrical fast
supply lines
transient/burst
IEC 61000-4-4
±1 kV for input/output
lines
±1 kV differential mode
Surge
IEC 61000-4-5
±2 kV common mode
<5 % U
(>95 % dip in U
0,5 cycle
40 % U
Voltage dips, short
(60 % dip in U
interruptions and
5 cycles
voltage variations on
power supply input
70 % U
lines IEC
(30 % dip in U
61000-4-11
25 cycles
<5 % U
(>95 % dip in U
5 sec
Power frequency
(50/60 Hz) magnetic
3 A/m
field IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Phoenix User's Manual
GUIDANCE AND MANUFACTURER'S DECLARATION –
ELECTROMAGNETIC IMMUNITY
Compliance level
±6 kV contact
±8 kV air
Not Applicable –
Battery powered
Not Applicable –
signal lines less
then 3 meters
Not Applicable –
Battery powered
T
) for
T
T
) for
T
Not Applicable –
Battery powered
T
) for
T
T
) for
T
3 A / m
Electromagnetic environment -
recommendations
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic material,
the relative humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Phoenix requires continued
operation during power mains interruptions,
it is recommended that the Empi Phoenix be
powered from an uninterruptible power supply
or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.