Hioki CM4371 Manual De Instrucciones página 21

Ocultar thumbs Ver también para CM4371:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Auto hold conditions
Display value updates are stopped when the following two conditions are satisfied:
• When the measured value exceeds the threshold value described in the table in the next page.
(voltage, current)
When the measured value is less than the threshold value described in the table in the next page.
(resistance, continuity, diode)
• When the range over which the measured value is fluctuating stabilizes within the fluctuation range
described in the table in the next page.
Held
measured
value
Measured
value
Start
If the measured value falls below the threshold value (voltage, current) or exceeds the threshold value
(resistance, continuity, diode) after display value updates are stopped, display value updates will stop once
more if the two conditions are satisfied again.
Auto hold
e.g.: 100.0 A
Fluctuation range
Fluctuation
range
End
End
Start
Measurement
Auto hold
e.g.: 99.0 A
Fluctuation range
Threshold value
e.g.: 12.0 A
Time
End
Start
Manual Hold / Auto Hold
17

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Cm4372Cm4373Cm4374

Tabla de contenido