Hioki CM4371 Manual De Instrucciones página 62

Ocultar thumbs Ver también para CM4371:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Accuracy Table
AC voltage (PEAK MAX/PEAK MIN)
(Accuracy guarantee
Range
range)
6.000 V
(0 V to ±12.00 V)
60.00 V
(±3.0 V to ±120.0 V)
600.0 V
(±30 V to ±1000 V)*
1000 V
(±50 V to ±1000 V)*
Maximum display count for all ranges: 1200/1700 counts
*1 Frequency range of 15 Hz≤f<20 Hz is designed value.
Within the frequency range of f<45 Hz, the accuracy guarantee assumes a superposed DC voltage of less
than 500 V.
*2 Values of up to ±1200 V are displayed, but accuracy is not defined for display values in excess of 1000 V
(which are provided as reference values).
*3 Values of up to ±1700 V are displayed, but accuracy is not defined for display values in excess of 1000 V
(which are provided as reference values).
58
Accuracy guarantee
Resolution
frequency range*
0.01 V
15 Hz≤f<45 Hz
45 Hz≤f≤66 Hz
66 Hz<f≤1 kHz
0.1 V
15 Hz≤f<45 Hz
45 Hz≤f≤66 Hz
66 Hz<f≤1 kHz
2
1 V
15 Hz≤f<45 Hz
45 Hz≤f≤66 Hz
66 Hz<f≤1 kHz
3
1 V
15 Hz≤f<45 Hz
45 Hz≤f≤66 Hz
66 Hz<f≤1 kHz
Measurement accuracy
1
±1.8% rdg.±0.07 V
±1.5% rdg.±0.07 V
±1.8% rdg.±0.07 V
±1.8% rdg.±0.7 V
±1.5% rdg.±0.7 V
±1.8% rdg.±0.7 V
±1.8% rdg.±7 V
±1.5% rdg.±7 V
±1.8% rdg.±7 V
±1.8% rdg.±7 V
±1.5% rdg.±7 V
±1.8% rdg.±7 V

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Cm4372Cm4373Cm4374

Tabla de contenido