How To Test Control Solution - TRUE METRIX GO Guía De Prueba Rápida

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

How To Test Control Solution

Expiration Dates
LOT - ABC1234
EXP - 2018/10/31
1 - 72-90 mg/dL
2 - 150-200 mg/dL
3 - 301-398 mg/dL
May 30, 2018
Test Strip Label
Control Solution Label
1. Gather and check supplies.
See Getting Started -Testing Checklist
(page 17).
2. Allow control solution,
3. Wash hands.
vial of test strips and
Dry thoroughly.
meter to adjust to room
temperature for 10
minutes.
20
5. Remove one
4. Gently swirl or
test strip from
invert control
solution bottle
vial. Close vial
to mix.
immediately.
DO NOT SHAKE.
Use test strip
quickly after
taking it out of
the vial.
6. Insert test strip into
7. Remove cap from
Test Port. Meter turns
control solution
on.
bottle. Gently
squeeze one drop
Keep test strip in meter
onto a clean tissue.
until testing is finished.
Wipe off bottle tip
Do not add control
and discard tissue.
solution to test strip
before inserting into
meter.
21