How to Disinfect the Meter
1. After cleaning the meter, use a fresh wipe to wet
all the outside surfaces of the meter.
~ Make sure no liquids enter the Test Port or
any other opening in the meter.
~ Do not spray meter with any cleaning agents.
Caution!
If the meter remains attached to the test strip vial
during cleaning, make sure that the vial cap is
completely closed before disinfecting.
2. Keep the meter wet for 2 minutes by rewiping all of the
surfaces. Fresh wipes may be used if needed to keep
the meter wet. Discard used wipes.
Caution!
Do not wrap or scrub the meter or the test strip vial
with the wipe to keep the surfaces wet. Wrapping wipe
around test strip vial or scrubbing can cause the label to
become unreadable.
3. Allow meter to air dry for an additional 1 minute.
48
4. Reattach meter to the new test strip vial by setting
the test strip vial on flat surface with vial lip facing to
the left.
~ With Test Port facing front, place bottom of meter firmly
on vial top. Meter must be seated flat on top of vial cap.
~ Holding the vial, twist the meter 1/4 turn clockwise.
The Test Port area on the meter should cover the vial lip
if attached properly.
5. Wash hands thoroughly after cleaning/
disinfecting the meter. Make sure your
hands are completely dry.
Caution!
If hands are not dry before removing a test
strip from the vial, the vial may become
contaminated. Test strips that have become
contaminated may result in incorrect results or
error messages.