SITE REQUIREMENTS
The COMPULAB 3 Dispensing System is designed to operate in most workplace conditions. Ensure that the work area
temperature range is 0°C to 40°C (32° to 104°F) and that the relative humidity remains between 10% and 90%. Locate
the dispensing system so that it is not exposed to running water and is protected from dangerous chemicals and
excessive or conductive dust.
POWER REQUIREMENTS
The COMPULAB 3 Dispensing System requires 115 (90–130) VAC or 230 (180–260) VAC, 50/60 Hz incoming power.
Each pump drive draws 1.7 A at 115 V or 0.8 A at 230 V.
INTERCONNECTION
Each pump drive is connected to the control console with a 3-1/2 ft (1.1 m) RJ-12 cable (catalog number 77095-02).
Longer cables are available as an option, but are not recommended for use in areas where interference may be high.
Use as short a cable as is practical. Use only cables supplied by Barnant Company.
Refer to Figure 4, and connect the components of the dispensing system. Ensure that the plugs snap firmly into the
connectors on the back of the control console and on the back of each pump drive.
POWER CONNECTION
Each pump drive requires AC power to operate. The power for the control console is supplied by one of the pump drives.
WARNING: The dispensing system must be plugged into a grounded power source only. An ungrounded power
source presents a serious shock hazard.
CAUTION: Ensure that the 115/230 voltage switch on the back of each pump drive is set to the appropriate
voltage before connecting the pumps to the source voltage. Selecting the incorrect voltage can
cause damage to the dispenser.
INSTALLATION
Figure 4. Interconnecting Components