Quick Start Guide
Button
Short Press
Long Press
Short Press
(Menu
Function)
CALIBRATION
When the balance is operated for the first time, a span calibration is
recommended to ensure accurate weighing results. With balance turned on:
Step-1
Press and hold Menu until [mMeNU] (Menu) is displayed.
Step-2
Release the button, the display will show [C.A.L].
Step-3
Press Yes to accept. [SpaN] will then be shown.
Step-4
Press Yes to begin the span calibration. Zero reading is stored
automatically.
Step-5
Place the specified span calibration mass or press No to select an
alternate weight.
[--C--] will be displayed while the reading is stored
The display will show [done] if the calibration was successful. The balance
returns to the previous application mode and is ready for use.
OPERATION & APPLICATIONS
The balance can be configured to operate in various Application modes. Press
and hold Mode key to select your Application mode.
The Scout SPX balance incorporates the following Applications:
[wWEIGH]
[Count]
Weighing
Counting
Ohaus Corporation www.ohaus.com
(866) 433-6682 • (281) 359-8538 • sales@novatech-usa.com • www.novatech-usa.com
Zero/On
Turns the
balance on;
If balance is
On, sets Zero
Off
Turns the
balance Off
Yes
Accepts the
current setting
[Percnt]
Parts
Percent
Weighing
Print
Initiates print
Initiates an
application
mode
Units
Changes the
Changes the
weighing unit
application
mode
No
Advances to
Moves to
the next menu
previous
or setting
menu item
[CHECk]
[total]
Check
Totalization
Weighing
iii
Scout
SPX Balances
Function
Tare
Mode
Enters the
User menu
Back
Exits the User
menu
[Hold]
Display
Hold
Scout
SPX Balances
Tare
Menu
Exit

Publicidad

Tabla de contenido
loading

Tabla de contenido