4. For each extension, lead, or plug, fully insert the torque wrench (packaged with the
neurostimulation system) into each self-sealing grommet of the connector block and
tighten each setscrew (Figure 2).
#
Cautions:
Be sure the torque wrench is fully inserted into the self-sealing grommet. If the
■
torque wrench is not fully inserted, the setscrew may be damaged, resulting in
intermittent or loss of stimulation.
Before tightening setscrews, ensure that the extension or lead connector pins are
■
inserted into the connector block to prevent damaging the lead or extension.
Verify that each leaf of the self-sealing grommet is closed after the torque wrench
■
is withdrawn. If fluid leaks through a grommet seal that is not fully closed, the
patient may experience shocking, burning, or irritation at the neurostimulator
implant location, intermittent stimulation, or loss of stimulation.
Figure 2. Tightening the setscrews in the self-sealing grommet.