RD SET™ MP Series
Patient Cables
Reusable
INDICATIONS
The RD SET™ MP Series patient cables and Masimo RD SET sensors are indicated for the continuous, noninvasive
monitoring of arterial saturation (SpO
PCX-2108A
02/13
DESCRIPTION
The RD SET MP Series patient cables are for use on Philips modules with Masimo SET® Technology and devices with
Philips FAST-SpO
Technology.
2
On Philips modules with Masimo SET Technology, RD SET MP Series patient cables are for use with RD SET disposable
sensors.
On devices with Philips FAST-SpO
sensors. Masimo SET Technology is not available on devices with Philips FAST-SpO
WARNING: Masimo sensors and cables are designed for use with devices containing Masimo SET® oximetry or licensed
to use Masimo sensors.
WARNINGS, CAUTIONS, AND NOTES
• Always refer to the oximeter module operator's manual for complete instructions or additional instructions.
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable
and discard if cracks or discoloration are found.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable
and sensor before use, otherwise degraded performance and/or patient injury can result.
• Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.
• Failure to properly connect the sensor or the oximeter module to the cable may result in intermittent readings,
inaccurate results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or
disconnecting either end.
• To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize
the cable or sensor.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may
damage the electrical components, potentially leading to patient harm.
INSTRUCTIONS
A) Attach the RD SET MP Series patient cable to the Device
1. Refer to Fig. 1. Orient the connector of the cable to mate with the patient cable connector on the Module.
Caution: Ensure you are connecting a RD SET MP patient cable to either a Masimo SET or Philips FAST-SpO
connector on the Module. There are a number of similar connectors with different colors and different mechanical
keying. Never force a patient cable connector into the module. Failure to use a RD SET MP patient cable may result in
damage to the module, inaccurate readings, or no readings.
Note: There is a difference in color shade between the RD SET MP Patient Cable and the Philips Module patient cable
connector. However, this is an acceptable configuration. Masimo SET performance is not available when connected
to a device with Philips FAST-SpO
B) Attaching the RD SET MP Series patient cable connector to an RD SET disposable sensor connector
1. Refer to Fig. 2. Orient the sensor connector to the patient cable connector as shown.
2. Refer to Fig. 3. Insert the sensor connector completely into the patient cable connector until it locks in place
as shown.
C) Disconnecting the RD SET MP Series patient cable connector from the RD SET disposable sensor connector
1. Refer to Fig. 4. Pull firmly on the sensor connector to remove it from the patient cable.
CLEANING
1. Remove the sensor from the patient and disconnect it from the patient cable.
2. Clean the surface of the cable by wiping it with a 70% isopropyl alcohol pad.
3. Wipe all surfaces of the cable.
4. Saturate another cloth or gauze pad with sterile or distilled water and wipe all surfaces of the cable.
5. Dry the cable by wiping all surfaces with a clean cloth or dry gauze pad.
CAUTION
• Do not immerse the connector on the cable in any liquid solution.
• Do not sterilize by irradiation, steam, autoclave, or ethylene oxide.
• Do not clean with chemicals not approved above.
DIRECTIONS FOR USE
Not made with natural rubber latex
LATEX
) and pulse rate for adult, pediatric, infant and neonatal patients.
2
Technology, RD SET MP Series patient cables are for use with RD SET disposable
2
Technology.
2
2
3
Non-Sterile
Technology.
9005B-eIFU-0915
en
cable
2