WelchAllyn ProXenon 350 Instrucciones De Uso Y Manual De Servicio página 28

Sistema de cámara montada en faro
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 74
24
Chapter 3
The ProXenon Headlight Camera System is intended for use in the electromagnetic environment
specified below. The customer or user of the ProXenon Headlight Camera System should assure
that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV
differential
IEC 61000-4-5
mode
±2 kV common
mode
Voltage dips,
>95% dip in 0.5
short
cycle
interruptions,
and voltage
60% dip in 5
variations on
cycles
power supply
input lines.
30% dip for 25
cycles
IEC 61000-4-11
>95% dip in 5
seconds
Power frequency
3 A/m
(50/60Hz)
magnetic field
IEC 61000-4-8
Welch Allyn ProXenon Headlight Camera System
Electromagnetic Immunity
Compliance
Electromagnetic Environment -
Level
Guidance
± 6 kV contact
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
± 8 kV air
synthetic material, the relative humidity
should be at least 30%.
±2 kV for power
Mains power quality should be that of a
supply lines
typical commercial or hospital
environment.
Not applicable
±1 kV
Mains power quality should be that of a
differential
typical commercial or hospital
mode
environment.
±2 kV common
mode
>95% dip in 0.5
Mains power quality should be that of a
cycle
typical commercial or hospital
environment. If the user of the ProXenon
60% dip in 5
Headlight Camera System requires
cycles
continued operation during power mains
interruptions, it is recommended that the
30% dip for 25
ProXenon Headlight Camera System be
cycles
powered from an uninterruptible power
supply or battery.
>95% dip in 5
seconds
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.