WelchAllyn ProXenon 350 Instrucciones De Uso Y Manual De Servicio página 29

Sistema de cámara montada en faro
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 74
Directions for Use
The ProXenon Headlight Camera System is intended for use in the electromagnetic environment
specified below. The customer or user of the ProXenon Headlight Camera System should assure
that it is used in such an environment.
Immunity
IEC 60601
Test
Test Level
Conducted
3 Vrms
RF
150 kHz to 80 MHz
IEC 61000-
4-6
Radiated RF
3 V/m
IEC 61000-
80 MHz to 2.5 GHz
4-3
Chapter 3 Care and Maintenance
Electromagnetic Immunity
Compli
Electromagnetic Environment - Guidance
ance
Level
Portable and mobile RF communications equipment
should be used no closer to any part of the ProXenon
Headlight Camera System, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
P
3 Vrms
d = (1.17)
P
3 V/m
d = (1.17)
80 MHz to 800 MHz
P
d = (2.33)
800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
b
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
25
a

Publicidad

Tabla de contenido
loading

Tabla de contenido