Operation
00:00 Logging [Log01]
A: V.Fib.Coarse
A: Sinus
Defib. Disabled
BP: 120 / 80 P : N 4
Activate Parox Disable
1. Using left/right navigation buttons (located to the left and right
side of the "OK" button) tab to the desired functional area of
the display. Once the desired area is highlighted, use the up/down
navigation buttons to scroll through to the desired selection.
2. When the desired functional area is highlighted, press OK to edit
the settings. (For example, to set the rhythm, highlight the ECG
functional area and push the OK button).
Note: While in the edit mode (point 2 above) the user may
press the shift key together with the OK key in order to stay in
the editing screen. This allows the user to make multiple changes
before returning to the main screen again.
3. When the selection is complete, press OK to implement
the new selection and return to the main display.
Note: To cancel the selection press C. This action returns to
the main display with the previous setting unchanged.
Setup: Remote Control Settings
Setup Menu
Set sensitivity for
Shake Sensor
shake sensor (1/5)
Display Backlight
Display Contrast
3
Autopower off
Com. Channel
Calibrate BP
Language
–
To activate the Setup Menu, press the Menu button.
Use the navigation buttons to select function. Use quick execution
buttons or the navigation buttons to make a selection.
Press OK to save.
Shake Sensor can be adjusted to set the sensitivity for detecting
the action of "checked responsiveness".
Setup Menu
Display Backlight
Shake Sensor
Display Backlight
OFF
Display Contrast
Autopower off
Com. Channel
Calibrate BP
Keeping the backlight
Language
off increases battery life
with more than 50%
Set ON
Display backlight can be turned on or off. Unless used in a dark
room, it is recommended to have the backlight off to preserve batt-
eries.
Note: The default setting is "off".
L.Lung: Normal
4
R.Lung: Normal
4
80
Heart:
Normal
4
Vocal:
Cough
4
M
Breathing Rate: 14
+
Set OFF
Setup Menu
Display Contrast (1-9)
Shake Sensor
Display Backlight
5
Display Contrast
Autopower off
Com. Channel
Calibrate BP
Language
Display Contrast can be adjusted between 9 (darker) and 1
(lighter).
Note: Default setting is 5.
Setup Menu
Operating Device
Shake Sensor
turns off after being
Display Backlight
inactive for (5 - 60):
Display Contrast
Autopower off
15 min
Com. Channel
Calibrate BP
Language
To preserve batteries, the Autopower off function turns off the
remote control if there has been no keyboard activity for a set
number of minutes decided by the user between 5–60 minutes. A
warning will be given one minute before the unit turns off.
Note: Default setting is 15 minutes.
Com. Channel
Setup Menu
Please connect to
Shake Sensor
Manikin via cable
Display Backlight
Display Contrast
Current channel: 1
Autopower off
Channel can only be
Com. Channel
changed when cable is
Calibrate BP
used.
Language
To be able to operate several Resusci Anne Simulator units in the
same vicinity, it is possible to select 5 different communication
channels.
Note: Communication channels can only be changed if the remote con-
trol device is connected via cable. Default setting is channel 1.
Setup Menu
Shake Sensor
Select Channel (1 - 5)
Display Backlight
Display Contrast
1
Autopower off
Com. Channel
Calibrate BP
Language
7
–
+
–
+
–
+
–
+