EMERGENCY INTERUPTION / TERMINATION OF TREATMENT
IF EMERGENCY INTERRUPTION / TERMINATION OF TREATMENT IS
REQUIRED OR IF LIQUID IS NOTED LEAK ING THROUGH THE CERVIX,
DO NOT IMMEDIATELY REMOVE THE B ALLOON FROM THE UTERUS.
If fitted, first, activate the red "Emergency Stop" button on the LCD
overlay which will activate the emergency routine and display
"TREATMENT STOP" message on the LCD and proceed to actively
withdraw the liquid from the balloon.
Turn off the power to the TCU. Then proceed to turn power back on to
the TCU. In doing so, the TCU will automatically recognise that the
"previous" treatment did not finish correctly. The TCU will display the
"TREATMENT STOP" message on the LCD and then proceed to actively
withdraw the liquid from the balloon.
REMOVE THE BALLOON FROM THE UTERUS ONLY AFTER
"FINISHED V: XX ml Withdraw Balloon" MESSAG E APPEARS ON
THE LCD SCREEN.
.
If the TCU encounters a technical problem during the procedure, it
will issue a "TREATMENT FAILED" message. REMOVE THE
BALLOON FROM THE UTERU S ONLY AFTER THE
"FINISHED V: XX ml Withdraw Balloon" MESSAGE APPEARS
16. After completion of the treatment cycle, the LCD will show the
following message:
If the TCU encounters a technical problem during the deflation
procedure, it will issue a "DEFLATION FAILED" message. DO NOT
REMOVE THE BALLOON FROM THE UTERUS. Instead, wait until
the device withdraws the liquid from the balloon. REMOVE THE
BALLOON FROM THE UTERUS ONLY AFTER THE
"FINISHED V: XX ml Withdraw Balloon" MESSAGE APPEARS
ON THE LCD SCREEN.
17. After 10 seconds of deflation time, the LCD will show the following
message
OR
TREATMENT STOP
Do NOT Remove
WARNING
ON THE LCD SCREEN.
FINISHING...
Do NOT Remove
WARNING
FINISHED V: XX ml
Withdraw Balloon
12