A) The Disposable Cartridge has been installed properly into the
TCU, and
B) No fluid is leaking out from the Disposable Cartridge, particularly
from the balloon.
In case of a leaking Disposable Cartridge that has not spilled
treatment fluid into the TCU, replace with a new one. USE EXTR A
CAUTION, AS THE METALLIC PART OF THE DISPOSABLE
CARTRIDGE MAY BE VERY HOT.
Turn the ―POWER SWITCH‖ back to the ―ON ‖ position. Wait until the
TCU reaches the ―READY FOR TREATMENT‖ state.
If treatment fluid has leaked into the TCU, it will need to be repaired.

PATIENT PREPARATION

1. Provide patient with adequate analgesia.
2. During and/or prior to the pre-heating of the treatment fluid, patient
preparation can be performed. Appropriate sterile technique for
vaginal/cervical preparation should be used.
3. Place patient in dorsal lithotomy position.
TREATMENT
After approximately 12 minutes, the LC D will show ‖READY FOR
TREATMENT‖ (see below ) and the TCU will beep, indicating that the
fluid has been heated up to treatment temperature (approximately
173°C) and the treatment procedure can be initiated. The ―Heating‖
LED turns OFF and the ―Ready‖ LED turns ON. The LCD Display
reads:
If the system is left unused, the treatment temperature will be
maintained for 35 minutes. After this period, the system will
automatically turn OFF. To reinitiate Warm-Up, turn the "POWER
SWITCH" to the "OFF" position, and restart the process.
The Disposable Cartridge must not be used if it has been heated
up and cooled down more than twice or exposed to ambient air
IF THE BALLOON IS NOT COMPLETELY DEFLATED AND
LIQUID AND / OR GAS IS SEEN IN BALLOON DURING
"READY FOR TREATMENT" STATE, DO NOT PROCEED
WITH THE TREATMENT. Instead, turn the "POWER
SWITCH" to the "OFF" position, disconnect the power cord,
substitute the Disposable Cartridge, and restart the process.
READY
FOR TREATMENT
NOTICE
for more than 2 hours.
WARNING
8