SimplyGo User Manual
Connecting the Optional Humidifier
Warnings
•
Only use the humidifier connector tubing supplied with the
SimplyGo Humidifier Connecting Tube kit (REF 1101172) or the
SimplyGo Humidifier Pouch and Tube kit (REF 1101602).
•
Do not overfill humidifier.
•
Do not reverse the oxygen input and output connections. Water from
the humidifier bottle will travel through the cannula back to the patient�
•
Do not use the humidifier while the device is in Pulse mode.
Note: Only connect the humidifier if prescribed� Only use the humidifier if the
device is in Continuous Flow mode.
To connect the humidifier:
1�
Attach the humidifier accessory pouch to the mobile cart (see the
instructions included with the pouch)�
2�
Remove the cap from the humidifier bottle�
3. Fill the humidifier with distilled water to the level indicated by the
manufacturer�
4�
Replace the cap and make sure it is tightened securely�
5�
Attach the humidifier bottle adapter tube to the humidifier bottle
by turning the wing nut on the bottle counterclockwise until it is
fastened securely�
6. Place the humidifier bottle assembly in the accessory pouch (REF
1083699)�
7�
Rotate the bottle so the humidifier adapter hose is away from the
body of the portable oxygen concentrator�
8�
Attach the oxygen tubing from the humidifier bottle assembly to the
oxygen outlet connector on the portable oxygen concentrator�
9�
Attach the cannula tubing to the humidifier bottle outlet�
10� After assembly is complete, make sure the oxygen is flowing through
the cannula�
Notes
•
A bubble humidifier may be used with the SimplyGo.
•
Application accessories shall include a means to reduce the
propagation of fire�
•
A Humidifier Accessory Pouch (REF 1083699) may be purchased to
enable use of a bubble humidifier�
28