Specifications; Operating Conditions; Continuous Mode - Philips RESPIRONICS SimplyGo Manual De Instrucciones

Ocultar thumbs Ver también para RESPIRONICS SimplyGo:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Specifications

Operating
Conditions
Storage and Transport
Conditions
SimplyGo Oxygen Concentrator
Oxygen
Concentration*
Flow Settings and
Pulse Volumes
Operational temperature:
41° F to 104° F (5° C to 40° C)
Relative humidity: 15% to 95%
Altitude: up to 10,000 ft� (3048 m�)
-4° F to 140° F (-20° C to 60° C) - unit only
(Note: Refer to the Battery Storage section of this
manual�)
Relative humidity: up to 95%, non condensing
87%-96% at all settings
Pulse Mode
1 = 12 ml; 1�5 = 18 ml; 2 = 24 ml; 2�5 = 30 ml;
3 = 36 ml; 3�5 = 42 ml; 4 = 48 ml; 4�5 = 54 ml;
5 = 60 ml; 5�5 = 66 ml; and 6 = 72 ml;
+/- 15% or 4 ml, whichever is greater (Average of 20
consecutive pulses)
up to a max of 2000 ml/min +/- 300 ml
Sleep Mode
Variable pulse volumes, based on breath rate, to
maintain a constant minute volume per setting�
1 = 250 ml; 1�5 = 375 ml; 2 = 500 ml; 2�5 = 625 ml;
3 = 750 ml; 3�5 = 875 ml; 4 = 1000 ml; 4�5 = 1125 ml;
5 = 1250 ml; 5�5 = 1375 ml; and 6 = 1500 ml
+/- 15% or 60 ml, whichever is greater (Sum of 20
consecutive pulses when the device is delivering at
a 20 pulses per minute rate)

Continuous Mode

�5 = 500 ml/min; 1 = 1000 ml/min;
1�5 = 1500 ml/min; 2 = 2000 ml/min;
+/- 15% or 150 ml/min, whichever is greater
(3 minute running average)
Note: Max recommended flow is 2 LPM (at nominal
outlet pressures of 0 and 7 kPa)�