Total hours used
83
JUNIOR MODE
If the patient will be using an Optiflow Junior nasal cannula (OPT316/OPT318), you must activate Junior
Mode. Do not use Junior mode for other patient interfaces.
Junior Mode limits the target settings to: 34 °C and 2 - 25 L/min, in increments of 1 L/min.
New target
settings
If you are unable to activate Junior Mode, it is possible that Junior Mode may not have been enabled for
your device. Contact your Fisher & Paykel Healthcare representative.
DAY/NIGHT MODES
You can set the myAIRVO 2 to "Day" mode or "Night" mode.
In "Night" mode, some of the myAIRVO 2 sounds will be made quieter.
The display will become dimmed. Alarms will be unaffected.
The myAIRVO 2 will remember its Day/Night setting when you switch it
off.
Press the Mode button to move on to the next screen.
COMPLIANCE
This screen displays three pieces of compliance data:
Total hours used
Hours per day
Checksum
Press the Mode button to return to the "Warm-up"/"Ready for use"
screen.
To activate Junior Mode:
You must be able to see the "Warm-up" symbol or the "Ready for use" symbol to activate
Junior Mode.
Hold the Mode button for 5 seconds.
New target settings
The target settings for dew-point temperature and flow will be changed automatically. The
colorful icons in the corners of the screen indicate that this unit is in Junior Mode.
To deactivate Junior Mode, follow the same procedure: hold the Mode button for 5 seconds.
To change the Day/Night setting:
Press the Up and Down buttons to choose the new
setting.
"Day"
à
Displays the total number of hours that the unit has
been switched on.
Displays the average number of hours that the unit has
been used per day.
Displays usage information for the medical practitioner.
A – 9
"Night"
à