SCHEDULE FOR CHANGING ACCESSORIES
The accessories for the unit must be changed frequently to avoid the risk of infection. Parts should be
replaced immediately if they are damaged or discolored; otherwise they must be replaced within the
periods shown in the following table. These periods assume that the correct daily and weekly cleaning
procedures and maintenance schedule described above are adhered to. If these procedures and schedules
are not followed, the maximum periods of use will change to those stated in the AIRVO 2 manual. These
accessories are for single-patient use only.
Maximum
period of
use
Optiflow Junior interfaces
1 week
OPT316
OPT318
All other patient interfaces
OPT942 / OPT942E
OPT944 / OPT944E
OPT946 / OPT946E
OPT970 / OPT970E
OPT980 / OPT980E
1 month
OPT842 / OPT842E
OPT844 / OPT844E
OPT846 / OPT846E
OPT870 / OPT870E
RT013 / RT013E
All tube & chamber kits
900PT561
900PT560 /900PT560E
2 months
900PT290E
900PT500 / 900PT500E
900PT501
900PT531
3 months
900PT913
or
1000 hours
Reusable
HC360
Some products may not be available in your country. Please contact your local Fisher and Paykel Healthcare representative.
FILTER REPLACEMENT
After the myAIRVO 2 has been switched on for a total of 1000 hours, a prompt will appear indicating that
an air filter change is due. Follow the steps below if filter change is due:
Air filter change
due
Replace now?
Now
Later
SERVICING
This device contains no internal serviceable parts.
Refer to the myAIRVO 2 Technical Manual for a list of external spare parts.
Part number and description
Nasal Cannula - Infant
Nasal Cannula - Pediatric
Optiflow™+ Nasal Cannula - Small
Optiflow™+ Nasal Cannula - Medium
Optiflow™+ Nasal Cannula - Large
Tracheostomy Interface
Mask Interface Adapter
Optiflow™ Nasal Cannula - Small
Optiflow™ Nasal Cannula - Medium
Optiflow™ Nasal Cannula - Large
Tracheostomy Interface
Mask Interface Adapter - 22mm
AirSpiral ™ heated breathing tube, MR290 auto-fill chamber and adapter
AirSpiral ™ heated breathing tube
MR290 auto-fill chamber and adapter
Heated breathing tube
Heated breathing tube, MR290 auto-fill chamber and adapter
Junior heated breathing tube, MR290 auto-fill chamber and adapter
(for use with OPT316/OPT318 only)
Air filter (or more often if significantly discolored)
Reusable water chamber
1.
Take the filter holder from the back of the unit and remove the filter.
2.
Replace the old filter with a new one.
3.
Reattach the filter holder to the unit (clip the bottom of the filter
holder in first, then rotate it upwards until the top clips into place).
4.
Press the Mode button to move on to the "Replace now" screen.
5.
Press the Up button to select "Now".
6.
Press the Mode button to confirm.
The hours counter will be reset to zero.
If you choose the "Later" option, the prompt will continue to appear
whenever the unit is switched on.
A – 14