PRIMARY SECONDARY
DISPLAYS
NING
CARTUCHO
(NO CASSETTE)
ER##
FALLA ELECTRONICA
(ERR## SYSTEM FAIL)
__
MINUTOS DE PAUSA
(HOLD MINUTES)
BLOQ
BOMBA BLOQUEADA
(LOCK PUMP LOCKED)
Page 50
CONDITION OR
PROBLEM
Cassette is not
installed or was
installed incorrectly
when user attempted
to run pump or prime
the feeding set.
Pump malfunction.
Hold time count
down. The primary
display shows the
remaining time.
Lock-out feature
disables FLUJO
(SET RATE), DOSIS
(SET DOSE), and
VOL CERO (CLEAR
VOL) dial settings,
as well as the pump's
prime features.
44838 Rev. J
EXPLANATION OR CORRECTIVE
ACTION
•
Turn dial to PAUSA (HOLD)
•
Insert cassette and close door
•
Turn dial to MARCHA (RUN) to begin
feeding, or attempt to prime the
feeding set again
•
Remove pump from patient
immediately and contact your local
Abbott Nutrition representative
•
Turn dial to OFF/CARGA (OFF/
CHRG)
•
Pump dial is at PAUSA (HOLD)
•
If desired, adjust the PAUSA (HOLD)
timer to any duration (in 1-minute
intervals) from 1 to 90 minutes by
depressing the Up and Down Arrow
Buttons
•
Turn dial to FLUJO (SET RATE) or
DOSIS (SET DOSE)
•
Depress both Up and Down Arrow
Buttons simultaneously to deactivate
lock-out feature
•
Repeat steps to activate