Guidance and manufacturer's declaration – electromagnetic immunity
The Abbott FreeGo pump is intended for use in the electromagnetic environment specified below. The customer or
the user of the Abbott FreeGo pump should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the Abbott FreeGo pump
is used exceeds the applicable RF compliance level above, the Abbott FreeGo pump should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the Abbott FreeGo pump.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Compliance
Level
Level
3 Vrms
3 V/m
44838 Rev. J
Electromagnetic Environment Guidance
Avoid using the Abbott FreeGo pump in close proximity
to electronic devices or equipment that emit radio-
frequency (RF) energy, such as X-ray and magnetic
resonance imaging equipment. Actively transmitting cell
phones and similar devices should be kept two or more
meters away from the pump.
Technical Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Abbott
FreeGo pump, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2√P
d = 1.2√P
d = 2.3√P
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
be less than the compliance level in each frequency
range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
80 MHz to 800 MHz
800 MHz to 2.5 GHz
should
a
Page 67