Guidance and manufacturer's declaration – electromagnetic immunity
The Abbott FreeGo pump is intended for use in the electromagnetic environment specified below. The customer or
the user of the Abbott FreeGo pump should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrostatic fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Page 66
IEC 60601 Test Level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV line(s) to line(s)
<5 % U
(>95 % dip in
T
U
) for 0.5 cycle
T
40 % U
(60 % dip in U
T
for 5 cycles
70 % U
(30 % dip in U
T
for 25 cycles
<5 % U
(>95 % dip in
T
U
) for 5 s
T
3 A/m
Compliance Level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV differential mode
<5 % U
(>95 % dip in
T
U
) for 0.5 cycle
T
)
40 % U
(60 % dip in U
T
T
for 5 cycles
)
70 % U
(30 % dip in U
T
T
for 25 cycles
<5 % U
(>95 % dip in
T
U
) for 5 s
T
Class 2 (3 A/m)
44838 Rev. J
Electromagnetic Environment
Guidance
For best protection from ESD, avoid
using the Abbott FreeGo pump in
environments that promote high
levels of static discharge (e.g.
synthetic floor materials in low
humidity environments).
Mains power quality should be
that of a typical hospital, acute
care, long term care, or residential
environment.
Mains power quality should be
that of a typical hospital, acute
care, long term care, or residential
environment.
Mains power quality should be
that of a typical hospital, acute
care, long term care, or residential
)
environment. The Abbott FreeGo
T
pump allows continued operation
during power mains interruptions via
)
the internal battery.
T
Power frequency magnetic fields
should be at levels characteristic
of a typical location in the typical
commercial or hospital environment.