Abbott FreeGo Manual De Uso página 54

Bomba para alimentación enteral
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

PRIMARY SECONDARY
DISPLAYS
FIN
DOSIS ADMINISTRADA
(DOSE DONE)
DIAL
MAL COLOCADO
(DIAL NOT SET)
CARG
POR FAVOR ESPERAR
(CHRG PLEASE WAIT)
BOMB
LISTA PARA USAR
(PUMP READY TO USE)
Page 52
CONDITION OR
PROBLEM
The pump has
completed delivering
the programmed
dose.
Pump dial is between
two dial positions.
For example: the
dial is stuck halfway
between MARCHA
(RUN) and PAUSA
(HOLD).
The battery charge is
below the minimum
acceptable level and
the pump cannot
be used until the
minimum charge
level is reached.
The battery charge
has reached the
minimum level and
the pump can now be
used.
44838 Rev. J
EXPLANATION OR CORRECTIVE
ACTION
If you are done feeding:
Turn dial to OFF/CARGA (OFF/
CHRG), the dose has been delivered.
This message is displayed because
VOL ADM (VOL FED) has reached
your DOSIS (SET DOSE)
Or... If there is more food to deliver:
Turn dial to VOL CERO (CLEAR
VOL) (after 5 seconds, volume will be
cleared)
Turn dial to DOSIS (SET DOSE),
and use Up and Down Arrow Buttons
to set your remaining dose to be
delivered
Turn dial to MARCHA (RUN) to begin
feeding
Turn the dial to the desired dial
position, directly aligning the two dots
(on the dial and the dial position label)
Continue setting up and operating
pump as normal
No action can be taken.
Pump will be available for use when
the display indicates "BOMB LISTA
PARA USAR" ("PUMP READY TO
USE")
Turn dial to any position to clear
message
Confirm all settings are accurate
Turn dial to MARCHA (RUN) to
continue feeding