CLEANING

For trouble-free operation, check the pump daily and clean immediately after spills
occur. Before cleaning, be sure that pump is turned OFF and UNPLUGGED.
Do not submerge, autoclave, heat, steam, ETO or radiation sterilize the pump.
• Clean outside surface only (do not spray water or cleaning solution
into pump).
General Cleaning Instructions:
• For general cleaning, use soft cloth and warm soapy water. Wipe and
dry thoroughly.
• Do not remove the rotor.
» After cleaning the rotor, verify that all rollers (on rotor) turn freely.
Disinfecting Instructions:
If necessary to disinfect pump, please follow the procedure below.
Extended exposure to cleaning solutions can cause damage to the pump.
• Suggested Disinfecting Solutions:
» 10% solution of 5.25% sodium hypochlorite (household bleach)
» 40% solution of ethyl alcohol
» 10% solution of phosphoric acid
» For additional disinfecting solutions - contact your local Abbott
representative
• Clean pump as instructed in general cleaning instructions
• Apply the disinfectant in the identified formulated mixture
• Let stand for 5 minutes minimum (or longer if specified by the manufacturer
of the solution)
• Wipe thoroughly with a damp cloth and air dry
NOTE: These recommendations are not substitutes for official procedures that may
differ among institutions.
44838 Rev. J