5. Overview

The Autester ST DRY PV are versatile and suitable autoclaves for a wide range of
applications, sanitation, industrial processes and quality control. These autoclaves
can sterilize solids, with or without being wrapped, liquids, culture media. They also
have disinfecting drying cycles.
5.1 Purge
The purging operation is performed by the autoclave to remove air inside the boiler
to fill the vapor space. Where there is no air steam.
The purge system must be suitable to sterilize intended load.
5.2 Atmospheric purge
In atmospheric vented comes out gravimetrically, impulsed by steam for a certain time.
Suitable for solid unwrapped (Tweezers, needles, scalpels ...).
5.3 Pulsed vacuum purge
A vacuum pump is used to remove air from the boiler. During one or more extractions
purged air is produced.
Suitable for wrapped loads, catheters, ...
5.4 Bowie Dick test and Vacuum test
Bowie Dick test used to verify operation of the purging by vacuum fractional.
The vacuum test designed to verify the operation of the vacuum pump and sealing
the autoclave for a certain time.
5.6 Sterilization not wrapped material
Place the instruments on waterproof paper and they do not touch each other.
5.7 Sterilization of wrapped materials and pipes
Only one instrument per bag.
Placing the bags upright they do not touch one another nor the walls of the auto-
clave.
The tubes must be clean, rinsed and drained. Ensure that the tubes do not form
folds that obstruct the inner diameter.
5.8 Sterilization of vessels
Never place sealed containers.
Place the containers upside down to prevent water tanks.
5.9 Sterilization of Liquids
Placing fluids in suitable containers at a temperature and above trays so that they
can collect spills.
The containers were filled about 2/3 full. Not close tightly, they will be plugged with
cotton or some sort of cap to facilitate exhaust of the pressure vessel and not be
formed.
Avoid using containers with narrow necks.
At the end of the sterilization cycle, the autoclave was let cool freely (until pressure
is 0 kg / cm²).
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(Sujetas a modificaciones sin previo aviso) Pag.: 21
Tel 34 93 770 08 77 Fax 34 93 770 23 62

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

4002420

Tabla de contenido