Alarms - J.P. SELECTA Autester ST Dry PV III Manual De Instrucciones

Autoclave con ciclos de vacío
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

INSTRUCTION MANUAL
Screen
ERROR 10
WATER RESERVOIR EMPTY
wait until pressure
and autoclave temperature
becomes low
and press OK
ACTUAL: +0.00BAR 23ºC
J
.P. SELECTA s.a.u.
CODE 80330 REV B
September 2017 (It can be modified without notice)

9. Alarms

If a failure occurs during operation, a message will appear. These
failures are indicated by showing ERROR on the screen:
OK
HH:MM:SS DD/MM/YY
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Error
Description
01
Error EEPROM faulty.
02
Pressure sensor error.
Error failure sensor or disconnected cable.
10
Water reservoir empty.
12
Lid / Door open.
13
The pressure has exceeded the maximum admissible
pressure of +2.5bar.
14
Vacuum error. Vacuum leackage or pump failure.
15
Pressure error during purge phase.
Insuficient water or water heater failure.
16
Overtemperature error. Internal temperature has
exceeded the maximum limit of 150ºC.
17
External heating element overtemp.
External 150ºC exceeded.
18
Sterilization error. Setpoint margins overcome or
there is air inside.
19
Error set value not reached. No water in the system
or heater failure.
20
Empty error, takes too long to empty the system.
Valves or drain blocked or failure.
21
Error no paper. Printer requires paper.
22
Printer error. Paper obstruction.
23
Power error, power failure (detected in the main
power source).
24
Vacuum test error.
Page: 30
Tel 34 93 770 08 77 Fax 34 93 770 23 62

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

4002420

Tabla de contenido