9. Alarms
If a failure occurs during operation, a message will appear. These
failures are indicated by showing ERROR on the screen:
OK
HH:MM:SS DD/MM/YY
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Error
Description
01
Error EEPROM faulty.
02
Pressure sensor error.
Error failure sensor or disconnected cable.
10
Water reservoir empty.
12
Lid / Door open.
13
The pressure has exceeded the maximum admissible
pressure of +2.5bar.
14
Vacuum error. Vacuum leackage or pump failure.
15
Pressure error during purge phase.
Insuficient water or water heater failure.
16
Overtemperature error. Internal temperature has
exceeded the maximum limit of 150ºC.
17
External heating element overtemp.
External 150ºC exceeded.
18
Sterilization error. Setpoint margins overcome or
there is air inside.
19
Error set value not reached. No water in the system
or heater failure.
20
Empty error, takes too long to empty the system.
Valves or drain blocked or failure.
21
Error no paper. Printer requires paper.
22
Printer error. Paper obstruction.
23
Power error, power failure (detected in the main
power source).
24
Vacuum test error.
Page: 30