MANUAL DE INSTRUCCIONES
************************
14/05/2015
09:52
========================
PROG: 0
MODE: SOLID2
TEMP: 134°C
TIME: 04'
DRY : 10'
========================
LOT: 000010
************************
TIME TEMP. PRESS. STATUS
023°C +0.0bar VACUUM
021°C -0.2bar VACUUM
021°C -0.4bar VACUUM
021°C -0.5bar VACUUM
022°C -0.5bar
HEAT
023°C -0.5bar
HEAT
023°C -0.5bar
HEAT
025°C -0.5bar
HEAT
060°C -0.3bar
HEAT
071°C -0.1bar
HEAT
080°C +0.0bar
HEAT
089°C +0.2bar
HEAT
095°C +0.3bar
HEAT
100°C +0.5bar
HEAT
095°C +0.0bar VACUUM
090°C -0.1bar VACUUM
086°C -0.3bar VACUUM
083°C -0.4bar VACUUM
081°C -0.4bar VACUUM
078°C -0.5bar VACUUM
076°C -0.5bar VACUUM
079°C -0.4bar
HEAT
088°C -0.2bar
HEAT
095°C +0.0bar
HEAT
101°C +0.1bar
HEAT
106°C +0.3bar
HEAT
110°C +0.5bar
HEAT
114°C +0.7bar
HEAT
119°C +0.9bar
HEAT
122°C +1.2bar
HEAT
126°C +1.4bar
HEAT
129°C +1.7bar
HEAT
133°C +2.0bar
HEAT
134°C +2.0bar
HEAT
03' 134°C +2.0bar STERIL
02' 134°C +2.1bar STERIL
01' 134°C +2.1bar STERIL
00' 134°C +2.1bar STERIL
132°C +1.9bar
EMPTY
124°C +1.1bar
EMPTY
112°C +0.4bar
EMPTY
104°C +0.1bar
EMPTY
00' 099°C +0.0bar
01' 093°C -0.2bar
02' 089°C -0.3bar
03' 085°C -0.4bar
04' 082°C -0.4bar
05' 079°C -0.5bar
06' 076°C -0.5bar
07' 073°C -0.6bar
08' 071°C -0.6bar
09' 069°C -0.6bar
068°C +0.0bar UNPRES
RESULT: PASS
J
.P. SELECTA s.a.u.
CODIGO 80330 REV B
8. USB data recorder
Device formatting type (USB pen drive):
Operation
Installation:
• Follow the instructions in the autoclave to activate the printer and the sending
interval of the readings.
Operation:
• By default the recorder is active with a recording interval of 1 minute. Follow
the instructions of the autoclave to modify the recording interval.
• Connect the «USB pen drive» to the number 6 connector in the equipment front.
Then, connect the «USB pen drive» to a computer to download the data file.
Important note: The «USB pen drive» must have a device name. It cannot
be blank.
The file created is text type, that can be opened with WordPad or WORD.
Sugested operation mode:
Record of an autoclave cycle needs about 3Kbytes.
An USB device 3GB memory can store millions of records.
Tip : Never erase any record of the USB device, you will record the complete life
of the autoclave.
DRY
DRY
DRY
DRY
DRY
DRY
DRY
DRY
DRY
DRY
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Septiembre 2017
(Sujetas a modificaciones sin previo aviso) Pag.: 29
FAT 32