Startup; Turn The Meter On And Off; Operation; Setup For Ph - Hach sensION+ Serie Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 40

Startup

Turn the meter on and off

N O T I C E
Make sure that the probe is connected to the meter before the meter is turned on.
Push and hold
to turn on or turn off the meter. If the meter does not
turn on, make sure that the batteries are properly installed.
Note: The meter is set to automatically power off after 5 minutes of inactivity to
maximize the battery life.

Operation

D A N G E R
Chemical exposure hazard. Obey laboratory safety procedures and
wear all of the personal protective equipment appropriate to the
chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
W A R N I N G
Pacemaker precautions. If electromagnetic interference (EMI) occurs,
it can either:
• Stop the stimulating pulses from the pacemaker that control the
rhythm of the heart.
• Cause the pacemaker to supply the pulses irregularly.
• Cause the pacemaker to ignore the rhythm of the heart and supply
pulses at a set interval.
Current research shows that cellular devices are not a significant
health problem for most pacemakers wearers. However, persons with
pacemakers should use precautions to make sure that their device
does not cause a problem. Keep the device a minimum of 20 cm
(7.9 in.) from the user.
10 English
C A U T I O N
Electromagnetic radiation hazard. Make sure that the antenna is kept at
a minimum distance of 20 cm (7.9 in.) from all personnel in normal use.
The antenna cannot be co-located or operated in conjunction with any
other antenna or transmitters.

Setup for pH

The meter is ready for pH or ORP calibration without additional setup.

Setup for dissolved oxygen

Before a calibration or a measurement for dissolved oxygen, the probe
must be polarized and the settings for atmospheric pressure and salinity
must be entered.
Polarize the electrode
If the probe or batteries are disconnected, connect the probe or install
the batteries and wait for polarization:
Disconnect time
< 5 minutes
5 to 15 minutes
> 15 minutes
Polarization time
10 minutes
45 minutes
6 hours