Place the tools upon the surfaces of the spherical shells. The
scanner will have contact first.
Carefully lift milling guide in the middle behind until the
milling tool also contacts the spherical shell surface.
Tighten the Allen screw on the milling side again.
34
7.
CalIBratIOn OF the
DevICe (aDJUstIng the
MagnIFICatIOn FaCtOr)
Before calibration, remove the vacuum hose from the table
and make sure that all four feet of the device rest upon the
lab workbench. The height of the feet can be adjusted by
turning.
7.1

at the gUIDe skID

CaUtIOn! Please do not loosen any screws at the device or
the radial milling guide except for those described herein.
Loosen screw CN of the vertical pantograph and adjust to the
desired value of the multiplier (indicated on the blank form).
Tighten screw again.