Tighten screw.
Reintroduce scanners into the holes as deeply as possible,
loosen the spherical shells using the foot switch and move
them horizontally. The scanners will completely slide into
their holes on their own. Fix the spherical shells again and
check the tools for smooth mobility.
Test calibration.
It is important that both tools can be inserted easily into the
calibration openings. Repeat calibration if necessary.
Upgrade kit
Multi-x
O P e r at I n g I n s t r U C t I O n s
8.
sPIrIt level
The spirit level is used for fine adjustment of the machine and
for selecting the insertion direction of abutments. The moun-
ting pegs of different size are used in accordance with the
individual implant system.
The spirit level is placed upon the respective mounting peg
and set into the abutment. Select a suitable peg so this has
some play within the abutment. The flat mounting of the spirit
level, the peg and the calibration plate (as well as the
abutment) guarantees vertical alignment of the components
in relation to each other.
gB
39