H. SPECIFICATIONS
PARI TREK
S PORTABLE AEROSOL SYSTEM SPECIFICATIONS
®
Weight: (Compressor only) ...................................................................................................................................................................0.8 lb
With Rechargeable Battery: ................................................................................................................................................................. 1.2 lbs
Dimensions: (Compressor only) ...................................................................................................................................4.5" x 3.5" x 1.75"
Certification: .............................................................................................................................IEC/EN 60601-1 and IEC/EN 60601-1-2
COMPRESSOR PERFORMANCE:
Maximum Pressure of Compressor: ................................................................................................................................. 35 PSI (2.4 Bar)
Maximum Liter Flow of Compressor: ............................................................................................................................................ 8.5 LPM
Operating Pressure of Compressor¹: ............................................................................................................................14.5 PSI (1.0 Bar)
Operating Liter Flow of Compressor¹: .......................................................................................................................................... 4.0 LPM
¹Measured with PARI LC
Sprint or LC
Plus Reusable Nebulizer attached.
®
®
AMBIENT OPERATION:
Temperature Range: ................................................................................................................................................................. 50ºF to 104ºF
TRANSPORT/STORAGE:
Ambient Temperature: ............................................................................................................................................................. 13ºF to 158ºF
Humidity: ........................................................................................................................................................................................ 10% to 95%
Atmospheric Pressure: .................................................................................................................................................. 500 hPa - 1060 hPa
Compressor Electrical Requirements: ....................................................................................................Input 12 VDC 1.67 A 18 watts
OPTIONAL RECHARGEABLE BATTERY:
Battery Type: .................................................................................................................................................................... Lithium Ion (Li-lon)
Operating time with battery: .......................................................................................................................... Approximately 40 minutes
Number of treatments per charge
...........................................................................................................Approximately 8 treatments
2
Normal time to charge: ......................................................................................................................................... Approximately 2 hours
2
Using a PARI Reusable Nebulizer to nebulize 2.5 ml of medication.
ELECTRICAL REQUIREMENTS:
Input: ......................................................................................................................................................................................... 12 VDC 1.67 A
Output: ..................................................................................................................................................................................... 12 VDC 1.67 A
MULTI-VOLTAGE AC ADAPTER:
FUHUA Model # UE24WU-120167SPA or equivalent
Input Voltage: ............................................................................................................................................... 100-240 VAC 50/60Hz 600mA
Output Voltage: ........................................................................................................................................................................12V DC 1.67A
EXPLANATION OF SYMBOLS
Attention : Read all 'Instructions for Use' before operation: ......................................................................................................
Insulation enclosed Class II equipment: ...........................................................................................................................................
Type "B" Protection from electrical shock: .....................................................................................................................................
Do not dispose of in unsorted municipal trash.
Please consult local laws and regulations for proper disposal: ...................................................................................................
Power Switch: .........................................................................................................................................................................................
'ON' Position (compressor running): ...............................................................................................................................................
'OFF' Position (compressor is not running): ...................................................................................................................................
NOTE: When switch is in the 'off ' position and the compressor is not running, there will still be electrical potential inside the
compressor housing. To remove power from the compressor, the power plug or battery must be removed.
EN-9