6. Hygienic preparation

This product may contain disposable items.
only once. So use these items only once and do
disposable items may impair the functionality and safety of the product and lead to
unforeseeable reactions as a result of ageing, embrittlement, wear, thermal load, the
effects of chemical processes, etc.

6.1 Intervals

You should check the filters and wipe clean the housing and the filter compartment
cover regularly. The headgear/headband can be washed as necessary. Please also
observe the following intervals.
Interval
Daily
Every 24
operating hours
Weekly
Every 1000
operating hours
Every 6 months
Annually
– Clean mask
– Clean hose system
– Clean bacteria filter in accordance with the operating
instructions
– Clean VENTIclick humidifier in accordance with the relevant
operating instructions.
– Clean the exhalation system after every use in accordance with
the operating instructions .
– Change the particle filter in the bacteria filter
– Clean the mask thoroughly in accordance with the operating
instructions.
– Clean coarse dust filter
– Change fine filter (filter change indicator
change it sooner
– Change coarse dust filter. Change it sooner if soiled or worn
– Change pressure measuring hose – sooner if soiled – (see "9.3
Change pressure measuring hose" on page 42).
– Change hose system.
Disposable items are intended to be used
reprocess them. Reprocessing
not
Activity
Hygienic preparation
). If it is soiled,