IPAP pressure range
EPAP pressure range
CPAP pressure range
Pressure accuracy
Increment
Minimum stable limit pressure
(PLS
) (min. pressure in event of
min
a fault)
Maximum stable limit pressure
(PLS
) (max. pressure in the
max
event of a fault)
Respiratory frequency
Accuracy
Increment
I:E (Ti/T):
Inspiration time
Increment
Accuracy
Trigger stage
Pressure increase speed
Pressure drop speed
Accuracy of volume measurement
Flow at max. speed at:
0 hPa
Tolerance
Flow at max. speed with bacteria
filter at:
0 hPa
Tolerance
Heating of respiratory air as per
HMV
Pressure constant measured to
EN ISO 17510 in CPAP mode
Fine filter separation level
to 2 µm
VENTImotion
(1 hPa = 1 mbar ≈ 1 cm H
20 % to 67 % of respiratory period
can be adjusted in 6 stages, separately for inspiration and expiration.
Trigger for expiration can be switched off in ST mode
300 l/min
±15 l/min
270 l/min
±15 l/min
2.5 ˚C
VENTImotion with VENTIclick
6 to 35 hPa
4 to 20 hPa
4 to 20 hPa
±0.6 hPa
0.2 hPa
≥ 0 hPa
≤ 60 hPa
6 to 45 1/min
± 0.5 1/min
1 1/min
1 %
±1 %
can be set in 6 stages
can be set in 6 stages
at 23 ˚C: ± 15 %
depending on heating stage
at 20 hPa: Δp ≤ 1 hPa
at 14 hPa: Δp ≤ 1 hPa
at 10 hPa: Δp ≤ 1 hPa
at 7 hPa: Δp ≤ 0.5 hPa
≥ 99.7 %
Technical specifications
O)
2
275 l/min
±15 l/min
250 l/min
±15 l/min
EN