7. Function check

7.1 Intervals

Perform a function check at least every 6 months. If you discover defects during the
function check, you must not use the VENTImotion again until these defects have
been rectified.

7.2 Method

1. Assemble the VENTImotion so that it is ready to function with the hose system,
exhalation system and the supply cable connection.
2. Close the opening of the hose system, for example using your thumb or hand.
3. Switch on the unit by pressing the on / off switch
4. If soft start is switched on, switch it off by pressing the soft start switch
Depending on the operating mode set, now check the following functions:
Function:
Pressure setting
Triggering
Soft start
Flow sensor
Alarms
If the values/functions set out below are not met/fulfilled, send the unit to your dealer
or to the manufacturer (Weinmann) for repair.
30
EN
Function check
/
Mode:
CPAP
S
T
ST
.
SX
SXX
.