Cleaning and disinfecting

This section updates the instructions for cleaning and disinfecting the ultrasound system, transducers, and
accessories.
Use the FUJIFILM SonoSite recommendations when cleaning or disinfecting your ultrasound system,
transducer, and accessories. Use the cleaning recommendations in the peripheral manufacturer's instructions
when cleaning or disinfecting your peripherals.
www.sonosite.com/products/transducers
See
Notes

Before getting started

Follow the disinfectant manufacturer's recommendations regarding appropriate personal protective
equipment (PPE), such as protective eyewear and gloves.
Inspect the system and transducer to determine that they are free of any unacceptable deterioration, such
as corrosion, discoloration, pitting, or cracked seals. If damage is evident, discontinue use, and contact
FUJIFILM SonoSite or your local representative.
Confirm that cleaning and disinfecting materials are appropriate for your facility's use. FUJIFILM SonoSite
tests cleaners and disinfectants for use with the FUJIFILM SonoSite systems and transducers.
Disinfectants and cleaning methods listed in this chapter are recommended by FUJIFILM SonoSite for
efficacy and material compatibility with the products.
Ensure that the disinfectant type and the solution strength and duration are appropriate for the equipment
and application.
Follow manufacturer recommendations and local regulations, when preparing, using and disposing of
chemicals.
WARNINGS
Cautions
8
The system and transducers must be cleaned and disinfected after each exam.
It is important to follow these cleaning and disinfecting instructions without
skipping any steps.
For cleaning the TEEx transducer, refer to the cleaning and disinfection
instructions in the TEEx User Guide.
Ensure that cleaning and disinfecting solutions and wipes are not expired.
Some cleaners and disinfectants can cause an allergic reaction to some
individuals.
Do not allow cleaning solution or disinfectant into the system connectors, or
transducer connector.
Do not use strong solvents such as thinner or benzene, or abrasive cleansers,
since these will damage the exterior surfaces. Use only FUJIFILM SonoSite
recommended cleaners or disinfectants.
for transducer images.