Approved cleaners/disinfectants for the transducer
Product
2
SaniCloth AF3
SaniCloth Plus
PI-Spray II
1
For maximum effectiveness, the component being cleaned must remain wet with disinfectant for a minimum
period of time.
2
Qualified for use as an intermediate-level disinfectant for mycobacteria.
3
The P11x transducer is not licensed for use in Canada.
4
The L52x transducer is for veterinary use only.
Refer to the cleaners and disinfection tool available at
more complete list of approved cleaners and disinfectants.
b Remove all gel, debris, and bodily fluids from the system.
c With a new wipe, clean the cable and transducer, starting from the cable, wiping toward the scanhead.
This method helps to avoid cross-contamination.
Caution
d Observe the minimum wet contact time recommended by the manufacturer. Monitor the transducer
for wet appearance. Re-apply with a new wipe if no longer wet.
8 Verify that all gel, debris, and bodily fluids have been removed from the system and transducer. If
necessary, repeat steps 5, 6, & 7 with a new wipe.
WARNING
12
Compatible Transducer
C11x, C60x, C60xi, C8x, HFL38x,
HFL50x, ICTx, L25x, L38x, L38xi, P10x,
3
, P21x, SLAx
P11x
4
HFL50x, L52x
C11x, C60x, C60xi, C8x, HFL38x,
HFL50x, ICTx, L25x, L38x, L38xi, L52x
3
P10x, P11x
, P21x, SLAx
Do not allow moisture near the electronic components of the connector.
Failure to remove all gel, debris, and bodily fluid could leave contaminants on the
probe.
4
,
www.sonosite.com/support/cleaners-disinfectants
Minimum wet contact time
3 minutes
3 minutes
10 minutes
Cleaning and disinfecting
1
for a